Equipment and Accessories for Modern High-Effective Technology of Ozone Therapy and Ozone prophylaxis
Equipment and Accessories for Modern High-Effective Technology of Ozone Therapy and Ozone prophylaxis
Equipment and Accessories for Modern High-Effective Technology of Ozone Therapy and Ozone prophylaxis
This website uses cookies to ensure
you get the best experience.
Ok
Leave feedback
The information which you give in completing this form will be forwarded to the designated party for its use and will not be provided by us to any other parties.

Oxygen Supply Equipment

The correctly chosen and properly operating Oxygen Supply System is one of the most mportant requirements for safe and efficient ozone therapy.

Oxygen gas can be supplied to Medical Ozone Generators of Medozons series from central oxygen network or oxygen tank. As oxygen source can also be used an oxygen concentrator or electrochemical generator of oxygen.
For proper function of Medical Ozone Generators of Medozons series the Oxygen Supply System should be equipped with an Oxygen Pressure Regulator adjustable from 0,3 to 1,5 bar (atm).
Oxygen Extra Purification Filter
Oxygen used in practice may contain micro-quantities of technical oils, moisture etc, thus extra purification of oxygen through a built-in carbon filter at the oxygen inlet port of ozone generator can be of great importance.

The filter used in Medical Ozone Generators of Medozons series ensures rough and fine purification of oxygen as well as its drying, and the ozonization chamber receives therefore guaranteed pure oxygen.
Autonomous Oxygen Supply System (as an example)
Designed for delivering medical oxygen of required quality to Medical Ozone Generators of Medozons series.
The system includes
2L flask, handle/stand, pressure reducer, flow regulator, manometer. The system is equipped with an adapter for oxygen filling from an external flask and a clamp for fixation of oxygen hose.
The system includes: 2L flask, handle/stand, pressure reducer, flow regulator, manometer. The system is equipped with an adapter for oxygen filling from an external flask and a clamp for fixation of oxygen hose.

The reducer is adjusted for proper work pressure of Medical Ozone Generators of Medozons series and constant oxygen flow-rate of 0,5 L/min and does not require any additional regulation.
Important Note!
Medical Ozone Generators of Medozons series supplied with oxygen from this system should be operated at ozone/oxygen flow-rate of 0,25 or 0,5 Liter per minute.
How to Fill the Autonomous Oxygen Supply System
In spite of simple filling procedure, it should be done by a technician familiar with the safety requirements for use of pressurized vessels.
First step
To fill the autonomous oxygen supply system use the adapter. For assembly use paronit insertions.
Second step
Remove the plug and fasten the adapter. Make sure that the valve of the system is closed.
Third step
Fasten the adapter onto a 40L oxygen flask, using a spanner tighten thread connections.
Fourth step
First open the valve of the oxygen system, then the valve of the big flask. Oxygen is conveyed from the flask to the oxygen system. The filling process is finished after whistle of "conveying" oxygen has stopped. Close the valve of the big flask, then the valve of the oxygen system. Disconnect the adapter.
Fifth step
Install the manometer (do not forget the insertion). The manometer shows residual oxygen pressure in the flask.
Sixth step
Connect the oxygen hose to the outlet of the oxygen system, use a PVC tube with diameter 6mm included into the connection set of each ozone generator of Medozons series, fix the connection with the clamp.
Seventh step
Connect the other end of the oxygen hose to the inlet of the ozone generator. The work place of ozone therapy practitioner is ready.
Eighth step
For oxygen delivery open the valve of the oxygen system.
Ninth step
The work place of ozone therapy practitioner, fitted out with the autonomous oxygen supply system.
Important Note!
The valve of the oxygen supply system should be opened for 2-3 minutes before work to make sure that the gas supply system of the ozone generator is filled with oxygen.
Important Note!
Rotation of colored parts of the reducer is forbidden because this can result in disturbances of regulation made by the manufacturer and failure of the unit.
Ozone Therapy
In Different Fields Of Medicine
Our news